Products in your basket: 0
Amount: €0.00
Online Shop » On-Line Measurement » Measurable Parameters » Sampler » Mobile & Stationary Units » X-000174
PB 150-SE/24 (Part No: X-000174)
Fully automatic stationary, self-emptying sampler for discontinuous sampling according to time, flow and events, for free standing installation.
Housing:
- Stainless steel (WKst. 4301) with 40mm hard foam insulation
- Separated wet section and control section
- Sample compartment temperature-controlled to 4 ... 6 °C with recirculation air
- Maintenance-free cooling unit with:
* CHC free refrigerant medium
* Automatic thawing device
* Heating
* 3-point control system with neutral zone
* Sun shield can be opened
- 4-point base
- Console or rack mounting.
Sampling system:
- Vacuum system with multifunction valve for the switching over of vacuum/pressure/ventilation
- Maintenance free and oil free vacuum pressure membrane pump with suction filter
- Suction power adjustable
* Medium suction speed >0.6m/sec at 5m suction height.
Dosage unit:
- Dosage vessel (Duran 50 glass)
- Conductive filling level electrodes in stainless steel (cond >50µS/cm)
- Motive pinch valve.
Dosing quantity: Single sample volume adjustable in the range 20 ... 350 mL
Control:
- Microprocessor C-MOS control
- Watertight foil keypad
- 4 x 20 digit
- Alphanumeric LCD display
- Key dialog operation
- Alternatively German or English
Program storage: 6 user programs
Program combinations: Time, flow, each combined with event
Time display: Date (WD, D, M, Y), Time (h, m, s)
Time preselection: Program start, Date/time with traceable bottle change
Time-dependent sampling: Interval in the range 0 ... 99:59 h
Sampling according to flow
- Analog input / 0 ... 999.9 samplings per hour at 20 A
- Digital input / pulse divider 1 ... 9999
Bottle filling: According to time, 0 ... 99:59 h / sample-dependent 0 ... 999
Filling limitation: Max. no. of samples per bottle No / Yes from 0 to 999
Program repetition: No / Yes (permanent operation)
Manual sampling: While program is running, with selection of bottle position
Flow signal input:
- Analog 0/4-20mA/ via optocoupler
- Threshold voltage 3.3 V/150 Ohm at 20 mA
- 2-point calibration at 4 mA and 20 mA
Quantity scaling: None/m3/h
Event input: Potential free closer (>50mS)
Distributor:
- Direct rotary distributor with start position recognition and adjustable automatic rinsing and cleaning cycle
- Distributor selection 12/24
- Automatic stop
Bottle rinsing:
- Pressure rinsing 2 to 4 bar
- Tapwater or very clean process water
- Rinsing duration 0 to 999.9 s
- Pause after rinsing 0 to 999.9 s
Bottle emptying:
- Automatic / manual
- Position preselection with tube pinch valve
- Bottle emptying 0 to 999.9 s
Bottles: 24 x 2.0 L glass (Duran 50)
Bottle positioning: Direct rotary distributor
Info memory:
- Max. 1000 entries
- Free selection of errors and events
- Graphic representation of 0/4-20 mA and temperature
Option: RS232 interface operation via plug connection (control unit housing) (in combination with cable and PC program only)
Option: Signal outputs / run and error message – program - program end - distributor
Sample pipeline:
- PVC
- Fiber-reinforced 1/2 inch (ID) with screw coupling
- Light-proof
- 5m long
Max. suction height / max. suction length: 7.5 m (from dosage vessel) / 30 m
Power supply:
- 230V
- 50Hz
- 2.5A
- Fuse 10A
- Shockproof socket
- In the case of permanent connection, with line separator
Pressure rinsing connection: Tapwater or very clean process water 2-4 bar R3/4 inch, (frost-proof installation)
Sample rinsing water outlet: 1 inch (ID)
Overvoltage protection: All inputs
Environmental conditions: Operation - 20... + 40 °C / storage: -25 ... +50 °C
Weight: Approx. 130 kg
Test certificates: CE
| Part Number | Part Description | Price | |
|---|---|---|---|
| X-000164 | PB 150-2/R24 Stationary Sampler | POA | |
| X-000165 | PB 150-2/R24G Stationary Sampler | POA | |
| X-000172 | PB 150-SE/12 | POA | |