50msec * Voltage on the control side 12V * Automatic pulse scaler (after specification of the max. number of samples per hour) Input for event signal: - Digital: Pulse input via potential-free closer contact * Pulse duration > 50msec * Voltage on the control side 12V * Automatic pulse scaler (after specification of the max. number of samples per hour) Signal output for rotary distributor: - 12 V DC Interfaces: - RS 232 - RS 485 - Centronics (parallel interface) Overvoltage protection of all inputs and outputs Power supply: - 230 V / 50 Hz - 10 A fuse protection - Power consumption approx. 250 VA Housing: - Stainless steel (Wkst. 1.4301) - Two parts cabinet design with arrangable sunshield for installation and maintenance activities - IP-65 type of protection (for control compartment) - Lockable door to control and dosing room with inspection window - Sample compartment with: * Double wall stainless steel enclosure * Temperature-controlled - Heat-insulated with separately lockable door and active, gentle, air recirculation. Environmental conditions: - Operation -20 ... +40 °C - Storage -25 ... +50 °C Weight: - Approx. 90 kg - depending on equipment Dimensions: - HxWxD: 1300mm x 710mm x 660mm (with open sun shield: height 1880mm) Test certificates: CE " /> PB 150-2/R12 Stationary Sampler (Part No: X-000162) - Mobile & Stationary Units > Sampler > Measurable Parameters > On-Line Measurement Products - Carl Stuart
 
Basket

Products in your basket: 0

Amount: €0.00

PB 150-2/R12 Stationary Sampler (Part No: X-000162)

Stationary sampler according to the vacuum principle

Fully automatic stationary sampler for discontinuous sampling according to time, flow or events, according to the vacuum principle; includes:
- Separated control and sample room (lockable)
- Suitable for installation in the open
- Independent temperature control
- Pinching valve with double tube pinching
- Adjustable pinching power
- Low-maintenance multifunction valve
- Automatic function monitoring of all valves
- Adjustable pump power
- Bottle filling in rows

Control:
- Microprocessor control
- Programmable in dialog operation
- 4 x 20 digit LCD display with background illumination
- Watertight foil keyboard
- Language selection

Dosage unit: Vacuum dosage unit
- Glass vessel of DURAN 50, 20 - 350 ml
- Motive pinching valve

Pneumatic:
- Oil-free vacuum pressure membrane pump with integrated filter
- Maintenance-free
- 12 VDC

Suction height: 7m from the dosage vessel

Pressure:
- Max. 1 bar (adjustable)
- Medium suction speed with ID ½ inch intake tube and 5m suction height > 0.6 m/sec
- Multifunction valve for switchover of vacuum/ ventilation/ pressure according to the tube pinching principle (like pinching valve)

Temperature control:
- CHC free cooling unit
- Frost protection
- Independent 3-point control system with neutral zone for cooling and heating of the sample compartment to +4 °C
- Overheating protection

Intake tube:
- PVC
- Fabric reinforced
- Screw fitting
- Suction hose ID ½"
- Length 5m

Sample container sets (selection with ordering):
- With sample distributor (from two containers)
- With carrying tray (from 12 sample bottles)
- Complete units exchangeable without problems
- 1 x 25 l PE collecting container
- 2 x 13 l PE collecting container
- 4 x 6.5 l PE sample bottles
- 12 x 2.9 l PE sample bottles
- 12 x 2.0 l glass bottles
- 24 x 1.1 l PE sample bottles
- 24 x 1.0 l glass bottles

Data storage for six user programs and status messages (operation and error messages)
- Single and combination programs (time - flow - event)
- Program repetition
- Program backup approx. 5 years

Time display:
- Hours, minutes, seconds, weekday, day, month, year

Preselection of time for instrument start:
- Date and time

Bottle filling time:
- 1 min....99 h, 59 min

Interval (filling pause):
- 1 min....99 h, 59 min

Overfilling protection:
- 1...9999 sample dosings per bottle

Input for flow signal:
- Analog 0/4-20 mA, load 50 Ohm
- Digital: Pulse input via potential-free closer contact
* Pulse duration > 50msec
* Voltage on the control side 12V
* Automatic pulse scaler (after specification of the max. number of samples per hour)

Input for event signal:
- Digital: Pulse input via potential-free closer contact
* Pulse duration > 50msec
* Voltage on the control side 12V
* Automatic pulse scaler (after specification of the max. number of samples per hour)

Signal output for rotary distributor:
- 12 V DC

Interfaces:
- RS 232
- RS 485
- Centronics (parallel interface)

Overvoltage protection of all inputs and outputs

Power supply:
- 230 V / 50 Hz
- 10 A fuse protection
- Power consumption approx. 250 VA

Housing:
- Stainless steel (Wkst. 1.4301)
- Two parts cabinet design with arrangable sunshield for installation and maintenance activities
- IP-65 type of protection (for control compartment)
- Lockable door to control and dosing room with inspection window
- Sample compartment with:
* Double wall stainless steel enclosure
* Temperature-controlled
- Heat-insulated with separately lockable door and active, gentle, air recirculation.

Environmental conditions:
- Operation -20 ... +40 °C
- Storage -25 ... +50 °C

Weight:
- Approx. 90 kg - depending on equipment

Dimensions:
- HxWxD: 1300mm x 710mm x 660mm (with open sun shield: height 1880mm)

Test certificates:
CE

Price: POA

 
Alternative Products
 
 Part NumberPart DescriptionPrice
 
 X-000160PB 150-2/R2 Stationary SamplerPOA
 X-000161PB 150-2/R4 Stationary SamplerPOA
 X-000163PB 150-2/R12G Stationary SamplerPOA